+86 19925884840 [email protected] +86 19925884840
, ,

50kW Z-Arm Digital Radiography System UMXRAY157 for General X-Ray Imaging

The UMXRAY157 50kW Z-Arm Digital Radiography System is a floor-standing digital X-ray system designed for general radiographic examinations in hospitals and diagnostic imaging facilities.

Equipped with a 50kW high-frequency X-ray generator, a 17 × 17-inch wired flat-panel detector, and an electronically adjustable Z-arm structure, the system supports flexible upright and table-based examinations. It features a 1000–1800 mm adjustable SID, a 9.5-inch LCD touch screen, and DR imaging software with DICOM workflow capabilities.

UMXRAY157 50kW Z-Arm Digital Radiography System

The UMXRAY157 50kW Z-Arm Digital Radiography System combines a 50kW high-frequency X-ray generator, a 17 × 17-inch wired flat-panel detector and an electronically adjustable Z-arm for general radiographic examinations. Its floor-standing arrangement supports upright and table-based positioning in hospitals and diagnostic imaging facilities.

The tube and detector move synchronously during positioning. The system offers electronic source-to-image distance (SID) adjustment from 1000 to 1800 mm, a 9.5-inch LCD touch screen and a DR workstation for patient registration, exposure control, image processing and DICOM-based workflows.

System Features

  • Flexible Z-arm positioning: Vertical travel, Z-arm rotation and independent tube/detector angle adjustment.
  • Electronic SID: Adjustable from 1000 to 1800 mm in vertical or horizontal configuration.
  • One-key positioning: Presets for flat-bed and chest radiography positions.
  • 17 × 17-inch wired detector: Amorphous-silicon panel with 3071 × 3071 pixel matrix and 16-bit quantization.
  • 50kW generator: 40–150 kV, 10–630 mA and 0.1–630 mAs specified ranges.
  • DR workflow: RIS worklist support, DICOM transfer to PACS, image processing, reports and examination templates.

Technical Specifications

Digital Flat Panel Detector

Detector type Wired amorphous-silicon flat-panel detector
Effective area 17 × 17 inches
Effective pixels Approximately 9.43 million
Pixel matrix 3071 × 3071
Pixel size 140 μm
Quantization depth 16 bit
Spatial resolution 3.7 lp/mm (manufacturer-listed)
Imaging time 7.6 s (manufacturer-listed)

Z-Arm Mechanical Structure

System structure Floor-standing Z-arm / scythe arm
Stand weight 210 kg
Whole-machine weight ≤ 288 kg
Vertical travel 600–1600 mm (±5%)
Z-arm rotation −45° to +135° (±2°)
Electronic SID adjustment 1000–1800 mm (±5%), vertical and horizontal
Tube assembly rotation −15° to +15° (±2°)
Detector rotation −45° to +45° (±2°)
Operator interface 9.5-inch LCD touch screen
Positioning presets One-key flat-bed and chest radiography positions
Minimum room height 2.4 m

High-Voltage Generator and X-Ray Tube

Generator power 50 kW
kV range 40–150 kV
mA range 10–630 mA
mAs range 0.1–630 mAs
Exposure time setting 1–6300 ms
Inverter frequency 60–200 kHz
Focal spot sizes 0.6 / 1.2 mm
Small / large focus power 20 / 50 kW
Maximum tube heat capacity 300 kHU
Maximum peak voltage 150 kV

Grid, Collimator and Radiography Table

Grid size 18 × 18 inches
Grid ratio ≥10:1
Grid mode Electrical mode (as listed)
Collimator indicator time 30 ± 3 s
Light field specification 150 W, 24 VAC (manufacturer-listed; confirm before procurement)
Radiation leakage ≤0.54 mGy/h (manufacturer-listed; test conditions not provided)
Table height 70 cm (±5 cm)
Table length 200 cm (±5 cm)
Table width 70 cm (±5 cm)
Table load capacity 135 kg

The specification sheet lists table and grid parameters but does not expressly define the full standard delivery scope. Confirm the table, grid and required accessories in the quotation.

DR Workstation and Imaging Workflow

Patient and Study Management

The workstation supports RIS worklist retrieval, manual examination registration and emergency quick registration. Operators can search historical studies, view images, prepare reports and transfer DICOM images to compatible PACS environments.

Exposure Control and Image Processing

Examination templates may load associated exposure settings, which can be adjusted before exposure. Other listed functions include focal-spot control, generator-status display, image acceptance or rejection, window/level adjustment, rotation, scaling, measurements and edge enhancement.

Film and Report Handling

The software supports image layout settings, film-printing queues, report output and configurable patient information display. Check interoperability with the receiving facility’s RIS/PACS and printing hardware.

Optional Functions

  • Automatic Exposure Control (AEC): Optional exposure-control function.
  • Image stitching: Optional image-combination feature.
  • EE-PACS: Optional PACS platform; separate from standard DICOM image-transfer functions.

These functions are optional and should be individually confirmed in a configuration-specific quotation.

Typical Applications

  • General radiographic examinations in hospitals and medical imaging departments.
  • Upright chest radiography and table-based examinations using the listed positioning presets.
  • Digital X-ray room installation or system upgrade, subject to room, power and regulatory assessment.

Installation and Procurement Considerations

The listed minimum room height is 2.4 m. The supplied documents do not provide a complete room footprint, electrical feed requirements or radiation-shielding design. Confirm room layout, shielding, electrical supply, commissioning, staff training and local regulatory requirements before installation.

For a quotation, provide the destination country, quantity, intended examinations, room dimensions, required table configuration, existing RIS/PACS details and desired optional functions.

Frequently Asked Questions

What flat-panel detector does the UMXRAY157 use?

The specifications list a 17 × 17-inch wired amorphous-silicon detector, 3071 × 3071 pixels and 16-bit quantization. No wireless detector option is specified.

Can the Z-arm adjust SID electronically?

Yes. The stated electronic SID adjustment range is 1000–1800 mm in vertical and horizontal configurations.

Are AEC and image stitching standard?

No. AEC, image stitching and EE-PACS are listed as optional functions.

What information is needed for a quotation?

Please provide the installation country, quantity, room dimensions, planned examinations, table and detector requirements, RIS/PACS requirements and desired options.

Related Radiology Equipment

Explore the UMY Medical Radiological Equipment collection, or view the UMDRXY005 portable digital X-ray system if portable imaging is required.

Request a Quote for the UMXRAY157

Share your installation country, room requirements, required configuration and quantity to request a tailored quotation.

Reviews

There are no reviews yet.

Be the first to review “50kW Z-Arm Digital Radiography System UMXRAY157 for General X-Ray Imaging”

Your email address will not be published. Required fields are marked *

Scroll to Top