UMIFIB004 Neonatal Infant Incubator
The UMIFIB004 is a neonatal infant incubator designed to provide a controlled thermal environment for routine newborn care in hospitals, maternity departments and neonatal care settings.
The system uses air-servo temperature control with an adjustable air temperature range of 20°C to 37°C. A transparent hood with multiple access ports allows clinical access and observation while maintaining the controlled incubator environment.
Audible and visual alarm functions monitor key operating conditions including power failure, fan failure, sensor failure and abnormal air temperature.
Key Features
Air Servo Temperature Control
Air-servo control maintains the selected incubator air temperature within a specified control range of 20°C to 37°C.
Stable Temperature Performance
Temperature fluctuation is specified at ≤0.5°C, with mattress temperature uniformity of ≤0.8°C under specified operating conditions.
Transparent Access Hood
The transparent dual-slope hood supports observation of the infant while providing multiple access points for routine care procedures.
Multiple Infant Access Ports
The hood configuration provides four operating ports, including two iris ports, together with tubing access grommets.
Audible and Visual Alarms
Safety monitoring includes alarms for power failure, fan failure, sensor failure and abnormal incubator air temperature.
Mobile Hospital Configuration
The incubator is mounted on a wheeled stand to support movement between appropriate neonatal and hospital care areas.
Applications
The UMIFIB004 infant incubator is intended for controlled newborn thermal-care environments where stable air temperature, continuous observation and convenient clinical access are required.
- Neonatal care units.
- Hospital newborn care.
- Maternity and obstetric departments.
- Neonatal observation workflows.
- Controlled thermal support for newborn care.
Technical Specifications
| Model | UMIFIB004 |
|---|---|
| Product Type | Neonatal Infant Incubator |
| Temperature Control Mode | Air servo control |
| Air Temperature Control Range | 20°C–37°C |
| Temperature Fluctuation | ≤0.5°C |
| Mattress Temperature Uniformity | ≤0.8°C |
| Temperature Rise Time | ≤45 min |
| Noise Level Inside Hood | ≤55 dB |
| Power Requirement | ≤400 VA |
| Mattress Tray Size | 630 × 360 mm |
Hood and Access Configuration
The incubator uses a transparent dual-slope hood designed to provide visibility together with convenient access for routine neonatal-care procedures.
| Mattress to Hood Height | 450 mm |
|---|---|
| Hood Configuration | Dual-slope observation hood |
| Operating Ports | 4 |
| Iris Ports | 2 |
| Tubing Grommets | 2 |
| Front Access Panel | 826 × 310 mm |
| Mattress Tray | 630 × 360 mm |
Safety Alarm Functions
The infant incubator provides audible and visual alarm functions for monitoring key operating conditions.
- Power failure alarm.
- Fan failure alarm.
- Sensor failure alarm.
- High air temperature alarm.
- Low air temperature alarm.
- Over-temperature alarm.
The matching platform specification lists a high-temperature deviation alarm at +3.0°C, a low-temperature deviation alarm at -3.0°C and an over-air-temperature threshold of ≤38.0°C.
Operating Conditions
| Environmental Temperature | 20°C–30°C |
|---|---|
| Relative Humidity | 30%–75% |
| Environmental Air Velocity | <0.3 m/s |
| Maximum Mattress Tray Load | 5 kg |
| Maximum IV Pole Load | 2 kg |
Temperature Control and Monitoring
The UMIFIB004 uses air-servo control to regulate the thermal environment inside the incubator hood. The system displays operating parameters including the selected temperature, measured air temperature and heating power status.
The specified air temperature control range is 20°C to 37°C. Temperature fluctuation and uniformity specifications should be considered together with the stated environmental operating conditions when planning installation and use.
Before Ordering
Please provide the following information so the required infant incubator configuration and project requirements can be confirmed before quotation.
- Required quantity.
- Destination country.
- Required power supply specification.
- Intended hospital or neonatal department.
- Required accessories.
- Whether an IV pole or additional accessories are required.
- Any installation or training requirements.
- Any tender, certification or documentation requirements.
Frequently Asked Questions
What is the UMIFIB004 infant incubator used for?
The UMIFIB004 is designed to provide a controlled thermal environment for newborn care in neonatal, maternity and hospital settings.
What temperature range does the infant incubator support?
The specified air temperature control range is 20°C to 37°C using air-servo temperature control.
How stable is the incubator temperature?
The specified temperature fluctuation is ≤0.5°C, while mattress temperature uniformity is specified at ≤0.8°C.
What safety alarms are provided?
The incubator provides audible and visual alarms for conditions including power failure, fan failure, sensor failure and abnormal air temperature.
How many access ports does the hood provide?
The matching hood configuration provides four operating ports, including two iris ports, together with two tubing grommets.
What information is required before quotation?
Please provide the quantity, destination country, required electrical specification, intended department and any accessory, installation or tender requirements.









Reviews